DFI INFINITY B
HIGH PERFORMANCE DIGITAL FORCE INDICATOR
USER’S GUIDE
www.cooperinstruments.com
PH: 540-349-4746 FAX: 540-347-4755
Przeglądanie stron 0
1 2 3 4 5 6 ... 89 90

Podsumowanie treści

Strona 1 - USER’S GUIDE

DFI INFINITY B HIGH PERFORMANCE DIGITAL FORCE INDICATOR USER’S GUIDE www.cooperinstruments.com PH: 540-34

Strona 2 - CONTENTS

Figure 4-1. Front Detail ITEM DESCRIPTION 1 or 6-digit, 9 segment, alphanumeric 0.68” high LED display with programmable decimal point. 2 SETP

Strona 3 - CF 156 iii M2544/N/0505

BUTTON Function in Configuration Mode Firmware Version Number and only after applying the power. Use ‘▲/MAX’ to select “MENU 2” or “MENU 1” the defau

Strona 4 - CF 156 iv M2544/N/0505

4.2 Rear of the Meter The following is a brief description of each part of the rear of the meter. The labeling on the top of the mounting sleeve (not

Strona 5 - CF 156 v M2544/N/0505

CONNECTOR # DESCRIPTION P1 AC Power Connector or DC Power Connector P2 External I/O Connector P3 Input Connector J4 Optional RS-232/RS-485 or Ethe

Strona 6 - 1.0 INTRODUCTION

Figure 5-1. Meter Exploded View Using Figure 5-1 as a guide, follow these simple instructions to open the unit: IMPORTANT: Turn-off the power and

Strona 7 - 2.0 UNPACKING

Figure 5-2. Board Assembly Removing/Installing Detail The meter is now disassembled to the point where you can check and configure jumpers and insta

Strona 8 - CF 156 3 M2544/N/0505

5.2.2 Optional Board(s) Installation To install optional printed circuit board(s): 1. “Reveal the Main Board” (refer to Section 5.2, Disassembly). 2.

Strona 9 - 4.0 PARTS OF THE METER

Figure 5-4. Signal Input Board Figure 5-5. Main Board S3A S3B (not installed) S3C Unlocks Pushbutton Programming Factory Calibration only, do not

Strona 10 - CF 156 5 M2544/N/0505

Figure 5-6. Ethernet Option Board Figure 5-7. RS-232/RS-485 Option Board JUMPER RS232 RS485 HALF DUPLEX RS485 FULL DUPLEXS1-A CLOSE OPEN OPEN S1

Strona 11 - CF 156 6 M2544/N/0505

Figure 5-8. 4 Relay Output Option Board The table below shows which jumpers are assigned to each relay. Defaults have asterisks. Table 5-3 4 Rela

Strona 12 - 4.2 Rear of the Meter

CONTENTS NOTES, WARNINGS AND CAUTIONS ... 11.0 INTRODUCTION ...

Strona 13 - 5.0 SETUP

assembly. Figure 5-10 shows the panel cutout dimensions, and the dimensions for the panel thickness. It requires that the following already be done:

Strona 14 - CF 156 9 M2544/N/0505

11. Set P1, the AC power connector, aside and connect or reconnect all other connectors to the back of the meter using Figure 4-3 in Section 4.2 as g

Strona 15 - CF 156 10 M2544/N/0505

6.3 VOLTAGE/PROCESS VALUE Input: 6.3.1 For ± 50mV, 0-100mV ranges: S1_K; S2_A/B/G/H, jumpers are closed as shown: NOTE: For some applications, the

Strona 16 - 5.2.3 How to Access Jumpers

6.5.2 For ±50mV, 0-100mV ranges with External Excitation: S1_L; S2_A/B/H, jumpers are closed as shown: 6.5.3 For ± 500mV, 1V ranges with Internal Ex

Strona 17 - CF 156 12 M2544/N/0505

6.5.10 For 0-20mA or 4-20mA range with External Excitation: S1_K; S2_A/B/C/D/H, jumpers are closed as shown with External Excitation: 6.5.11 POTENT

Strona 18 - CF 156 13 M2544/N/0505

Figure 7-2. Current Input with Sensor Excitation Figure 7-3. Voltage Input Without Sensor Excitation Figure 7-4. 3-Wire Voltage Input With Sensor

Strona 19 - 5.2.4 Panel-Mount Assembly

Figure 7-6. Potentiometer Connections with Internal Power Supply and Ratio Measurement. Figure 7-7. Potentiometer Connections with External Power

Strona 20 - CF 156 15 M2544/N/0505

Figure 7-10. RTD Connections with 3-wire RTD Figure 7-11. RTD Connections with 4-wire RTD 7.4 Signal Input Connections - Strain The following Fig

Strona 21 - 6.0 JUMPER SETTINGS

Figure 7-14. Voltage Input Without Sensor Excitation Figure 7-15. 3-Wire Voltage Input With Sensor Excitation Figure 7-16. 4-Wire Voltage/Bridge

Strona 22 - CF 156 17 M2544/N/0505

Figure 7-18. Potentiometer Connections with Internal Power Supply and Ratio Measurement Figure 7-19. Potentiometer Connections with External Power

Strona 23 - CF 156 18 M2544/N/0505

7.5 Cold Junction Compensation Board Installation for Thermocouples ... 247.5.1 Cold Junction Compensation Board Wi

Strona 24 - 7.1 Introduction

7.5.1 Cold Junction Compensation Board Wiring Figure 7-21 shows the wiring hookup for thermocouples. Use the following table to determine which color

Strona 25 - CF 156 20 M2544/N/0505

FRANCE GERMANY T/C TYPE WIRE COLORS WIRE COLORS + LEAD - LEAD + LEAD - LEAD J Yellow Black Red Blue K Yellow Purple Red Green T Yellow Blu

Strona 26 - CF 156 21 M2544/N/0505

USA WIRING CODE EUROPE WIRING CODE CONNECTION P1 PIN # ON ORANGE CONNECTOR Black White Green Brown Blue Green/Yellow ~ AC Line (L) ~ AC Neutral (N)

Strona 27 - CF 156 22 M2544/N/0505

8.0 SETPOINT SETTINGS MENU FLOWCHART Once this ‘SETPTS’ button is pressed, each of the four setpoints’ value will be displayed for approximately 10

Strona 28 - CF 156 23 M2544/N/0505

"SP 3", "SP3’s value" "SP 4", "SP4’s value" "RUN". Press ‘▲/MAX’ 4) Enables the user to select and

Strona 29 - CF 156 24 M2544/N/0505

10.0 MAIN MENU CONFIGURATION FLOWCHART 1) Once lockout security ID code is set “enable”, the Menu will not be displayed through Main Configuration M

Strona 30 - CF 156 25 M2544/N/0505

11.0 INPUT TYPE MENU FLOWCHART: (INPUT) 12.0 INPUT TYPE SELECTION: (INPUT) 12.1 Input Type: (Thermocouple) INPUt ENTER INPUT TYPE MENU: Press ‘MEN

Strona 31 - 7.6 Connecting Main AC Power

below for types). If the display matches your choice of Thermocouple. Go to Step 8. Press ‘▲/MAX’ 7) Scroll through the available thermocouple types

Strona 32 - 7.7 Connecting Main DC Power

VOLt “VOLt” DC VOLTAGE SUBMENU: Press ‘▲/MAX’ 4) Scroll through the available selection to "VOLt" (flashing). Press ‘RESET/ENTER’ 5) Disp

Strona 33 - CF 156 28 M2544/N/0505

Press ‘MENU’ 1) Display shows "INPUt" Input Menu. Press ‘RESET/ENTER’ 2) Display flashes "VOLt" (default) or previous stored inp

Strona 34 - CF 156 29 M2544/N/0505

25.0 COMMUNICATION CONFIGURATION FLOWCHART ... 5926.0 COMMUNICATION MENU CONFIGURATION: (COMM) ...

Strona 35 - CF 156 30 M2544/N/0505

13.0 READING CONFIGURATION MENU AND FLOWCHART: (RDG.CNF) CF 156 35 M2544/N/0505

Strona 36 - CF 156 31 M2544/N/0505

14.0 READING CONFIGURATION MENU: RdG.CNF ENTER READING CONFIGURATION MENU: Press ‘MENU’ 1) Twice though Main Menu items and display shows "RdG

Strona 37 - 12.2 Input Type: (RTD )

OFFSET MENU CONFIGURATION. Enter READING SCALE & OFFSET FORMAT SUBMENU: Press ‘RESET/ENTER’ 7) To enter Reading Scale & Offset Format Confi

Strona 38 - 12.5 Input Type: (Bridge)

Press ‘RESET/ENTER’ 22) Display shows "xxxxxx" some 6-digit number, represents the value of "INPUt 1" as taken from live input so

Strona 39 - CF 156 34 M2544/N/0505

Press ‘RESET/ENTER’ 5) Display shows "dP.ActV" Decimal Point Effect Selection menu. Press ‘RESET/ENTER’ 6) Display flashes "yES"

Strona 40 - CF 156 35 M2544/N/0505

14.5 FLS.dSP Enter FLASHING DISPLAY CONTROL Submenu: Press ‘RESET/ENTER’ 23) Display shows "SP1.FLS" Setpoint 1 Flashing. Press ‘RESE

Strona 41 - CF 156 36 M2544/N/0505

15.0 INPUT CONFIGURATION FLOWCHART CF 156 41 M2544/N/0505

Strona 42 - CF 156 37 M2544/N/0505

16.0 INPUT CONFIGURATION MENU: (INP.CNF) INP.CNF ENTER INPUT CONFIGURATION MENU: Press ‘MENU’ 1) Three times though Main Menu items and display

Strona 43 - CF 156 38 M2544/N/0505

In case of "tC" input type: Press ‘RESET/ENTER’ 10) Display shows "tC.CO.JC" Cold-Junction Compensation for "tC" The

Strona 44 - CF 156 39 M2544/N/0505

Adjustment. *The following Step 17 is a safety measure to prevent unintentional calibration. Press ‘▲/MAX’ 17) Display still shows "CAL 0"

Strona 45 - CF 156 40 M2544/N/0505

CF 156 v M2544/N/0505 32.1.12 Flashing “dP OVF” (Decimal Point Overflow) ... 7332.2 T

Strona 46 - CF 156 41 M2544/N/0505

ON.L INE Press ‘RESET/ENTER’ 30) Display flashes "ON.L INE" On line method. Press ‘▲/MAX’ 31) Scroll between 2 available selections: "

Strona 47 - CF 156 42 M2544/N/0505

Input Signal Range Scale Input 1 Value Scale Input 2 Value 0-50mV 000000 050000 0-100mV 000000 100000 0-1V 000000 100000 0-5V 000000 050000 1-5V 01000

Strona 48 - CF 156 43 M2544/N/0505

17.0 MULTI-POINT SCALE & OFFSET MENU CONFIGURATION FLOWCHART 18.0 MULTI-POINT SCALE & OFFSET MENU CONFIGURATION: (MP.SC.OF) For proper Mul

Strona 49 - CF 156 44 M2544/N/0505

This scaling method based on 2 Input values entered with 2 corresponding readings. Up to 10 Linearization Points can be selected to customize the tra

Strona 50 - CF 156 45 M2544/N/0505

To scale "MANUAL", calculate inputs based on transducer specifications and manually enter them via on the front panel push buttons. The foll

Strona 51 - CF 156 46 M2544/N/0505

READING/INPUT SCALE & OFFSET MODE DISPLAY READING VALUES (y = ax + b) MP.SC.OF IN.SC.OF Rd.SC.OF Yrd = (SCALE).Xin + (OFFSET) disable disable en

Strona 52 - CF 156 47 M2544/N/0505

OUt.CNF Enter OUTPUT CONFIGURATION MENU: Press ‘MENU’ 1) 5 times or scroll though main Menu until Display shows "OUt.CNF" Output Configu

Strona 53 - CF 156 48 M2544/N/0505

20.2 Ot.SC.OF Enter OUTPUT SCALE & OFFSET CONFIGURATION Submenu: READING 1 Press ‘RESET/ENTER’ 10) Display shows "REAd 1" Reading

Strona 54 - CF 156 49 M2544/N/0505

22.0 SETPOINT CONFIGURATION MENU: (SP.CNF) Setpoints 1 through 4 can be configured for a very large variety of zone and level signaling. SP1 and SP2

Strona 55 - CF 156 50 M2544/N/0505

23.0 ALARM CONFIGURATION MENU FLOWCHART 24.0 ALARM CONFIGURATION MENU: (AL.CNFG) SP3 and SP4 are often used as Alarm1 and Alarm2, because they have

Strona 56 - 20.1 ANA.OUt

NOTES, WARNINGS AND CAUTIONS Information that is especially important to note is identified by these labels: • NOTE • WARNING • CAUTION • IMPORTANT N

Strona 57 - CF 156 52 M2544/N/0505

Alarm configuration is used to select or setup: • The Active Zone for each alarm point (SP3 or AL1, SP4 or AL2) to above or below the setting with &q

Strona 58 - CF 156 53 M2544/N/0505

ACt IVE Press ‘RESET/ENTER’ 5) Display shows "ACt IVE" Alarm Active State submenu. Press ‘RESET/ENTER’ 6) Display flashes "AbOVE&qu

Strona 59 - CF 156 54 M2544/N/0505

There are 4 types of Alarm modes/Deviation Functions which can alter the alarm response as flowing illustrations: All horizontal axes represent time

Strona 60 - CF 156 55 M2544/N/0505

Press ‘RESET/ENTER’ 15) Display flashes "PROC." Process (No Deviation) mode or previous selection. Press ‘▲/MAX’ 16) Scroll 4 available fo

Strona 61 - CF 156 56 M2544/N/0505

25.0 COMMUNICATION CONFIGURATION FLOWCHART 26.0 COMMUNICATION MENU CONFIGURATION: (COMM) Communication Operation Requirement: Communication Option

Strona 62 - CF 156 57 M2544/N/0505

Purchasing the controller with Serial Communications permits an instrument or more meters to be configured or monitored remotely from an IBM PC compat

Strona 63 - CF 156 58 M2544/N/0505

Press ‘RESET/ENTER’ 10) Display shows "StOREd" stored message momentarily and then advances to the "StOP" Stop Bit setting submen

Strona 64 - CF 156 59 M2544/N/0505

ECHO Press ‘RESET/ENTER’ 20) Display shows "StOREd" stored message momentarily and then advances to the "ECHO" Line Feed setting

Strona 65 - CF 156 60 M2544/N/0505

• Message Hand Shake: The RTS (Request To Send) line from the host controller is checked when the device is ready to send measurement data. If the RT

Strona 66 - CF 156 61 M2544/N/0505

button. • For Universal meters, there is a list of the transmitted character (Hex number value) corresponding for each of the sixteen (16) possible S

Strona 67 - CF 156 62 M2544/N/0505

Choose between Ethernet or Serial Communication option per meter. Serial communications come with one 6 ft. communications cable with phone plug term

Strona 68 - CF 156 63 M2544/N/0505

SEPARA Press ‘RESET/ENTER’ 51) Display shows "StOREd" stored message momentarily and then advances to the "SEPARA" Separator S

Strona 69 - CF 156 64 M2544/N/0505

27.0 DISPLAY COLOR SELECTION MENU FLOWCHART: 28.0 DISPLAY COLOR SELECTION MENU: (COLOR) COLOR Enter COLOR SELECTION MENU: Press ‘MENU’ 1) Nine tim

Strona 70 - CF 156 65 M2544/N/0505

28.1 N.COLOR Enter NORMAL COLOR SUBMENU Press ‘RESET/ENTER’ 2) Display shows "N.COLOR" Normal Color submenu that allows user to change norm

Strona 71 - CF 156 66 M2544/N/0505

29.0 LOCKOUT (ACCESS SECURITY) CONFIGURATION MENU FLOWCHART: CF 156 68 M2544/N/0505

Strona 72 - 28.5 AL2.CLR

30.0 LOCKOUT (ACCESS SECURITY) CONFIGURATION MENU • To prevent unauthorized tampering with the setup parameters, this instrument provides protection

Strona 73 - CF 156 68 M2544/N/0505

The meter case label gives the names (abbreviated functions) of each of the twenty pins of P2, the center-bottom connector. Refer to Figure 30-1. F

Strona 74 - 31.0 EXTERNAL CONTROL LINES

31.8 +5 V (PIN 8) Up to 20 mA is available for driving external devices, but isolation should be provided if there is a possibility of common mode (gr

Strona 75 - CF 156 70 M2544/N/0505

Figure 30-2 For Setpoint Transistors 32.0 TROUBLESHOOTING - DISPLAY MESSAGES AND TROUBLESHOOTING GUIDE A flashing alpha-numeric message in the disp

Strona 76 - CF 156 71 M2544/N/0505

32.1.9 Flashing “R OVSC” (Reading Overscale) This display occurs when the reading scale and/or offset applied to the input signal causes the display t

Strona 77 - CF 156 72 M2544/N/0505

broken or otherwise opened. Check thermocouple or RTD wiring. “I OVSC” POSSIBLE CAUSE: TO CORRECT: The input scale and/or offset values chosen are l

Strona 78 - 32.2 Troubleshooting Guide

When you ordered your meter, you will receive the following items in the shipping box: QTY DESCRIPTION 1 Unit in a Mounting Sleeve with Gasket 1

Strona 79 - 33.4 Thermocouple Input

For accurate full scale measurements below -40 degrees, you may need to complete the full scale cold junction calibration procedure (Refer to Section

Strona 80 - 33.6 General

resolution; 0.1% accuracy; 6 msec step response. RELAY OUTPUT, OPTIONAL: Form -C, SPDT Relays POWER RATING FOR RESISTIVE LOADS: Two relays at P6 a

Strona 81 - CF 156 76 M2544/N/0505

1. Test with Analog Out (AN03) Sample Rate Process (100 mV) Rate (ms) TC (K type) Rate (ms) 0 7(135) 7(135) 1 14(68) 14(68) 2 27(36) 27(36) 3 52(

Strona 82 - CF 156 77 M2544/N/0505

Figure 33-1. Meter Housing and Panel Cutout 33.7 Flashing Messages: DURING PROGRAMMING (CONFIGURATION MODE) NUMERICAL OVERFLOW: “999999” OFFSET OVE

Strona 83 - 33.7 Flashing Messages:

DURING SETPOINT ADJUST (RUN MODE) OUT OF SELECTED DIGIT RANGE: “999999” NOT STORED IN EEPROM: “NOSTOR” VALUE PUT IN EEPROM: “STORED” 33.8 ITS-90 T

Strona 84 - 3.10 ITS-90 RTD 385 Table

33.11 ITS-90 RTD 392 US Table Temp oC Resistance Temp oC Resistance Temp oC Resistance -200 17.0787 210 180.9644 620 324.1198 -190 21.4575 220 18

Strona 85 - 34.0 FACTORY PRESET VALUES

INPUT SELECTION : INPUt VOLTAGE: VOLt Voltage type = UNIPOL Range = 100MV If choose bI POL Range = 50MV If choose different INPUt-- If tC: Type = J tC

Strona 86 - CF 156 81 M2544/N/0505

LOCKOUT CONFIG : LCk.CNF (ALL DISABLED) SP1 SP2 SP3 SP4 INPUt RdG.CNF INP.CNF MP.SC.OF OUt.CNF SP.CNF AL.CNF COMM COLOR 35.0 RECORD YOUR SETUP VAL

Strona 87 - CF 156 82 M2544/N/0505

ALARM CONFIGURATION : AL.CNF AL.CNFG = If ENAbLE: AL1.CNF ACtIVE = LAtCH = N.StAtE = AL.MOdE AL2.CNF: ACtIVE = LAtCH = N.StAtE = AL.MOdE AL db = NU

Strona 88 - CE APPROVALS INFORMATION

Electrical Safety EN61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory. Double Insulation Pollution De

Strona 89 - WARRANTY REPAIR POLICY

3.0 SAFETY CONSIDERATIONS 3.1 Safety Considerations This device is marked with the international caution symbol. It is important to read this manua

Strona 90 - CF 156 85 M2544/N/0505

CF 156 85 M2544/N/0505 Advance authorization is required prior to the return to Cooper Instruments. Before returning the item, contact the Repair D