CF 28 1 4/2001
DCM 460 VOLTAGE BRIDGE AMPLIFIER
Description
The DCM 460 Bridgesensor is a self-contained, AC powered, signal conditioning module for bridge type
instrumentation. It contains a precision differential instrumentation amplifier with filtered output and a highly
regulated, low noise, adjustable output bridge excitation source. The unit is completely encapsulated for use in
rugged environments. The output filter cutoff frequency is 2000Hz and may be lowered by adding two external
capacitors.
Features
2 kHz Frequency Response
Rugged, Compact And Fully Encapsulated
Complete System-just Add AC Power
Ready To Use With Screwdriver Wiring
Stable and Accurate
Applications
Weighing With Load Cells
Long Term Structural Monitoring
Process Control Pressure Transducers
Temperature Controls Using Thermocouples
Strain Measurements
Figure 1. Simplified Block Diagram
Przeglądanie stron 0
1 2 3 4 5 6

Podsumowanie treści

Strona 1 - CF 28 1 4/2001

CF 28 1 4/2001 DCM 460 VOLTAGE BRIDGE AMPLIFIER Description The DCM 460 Bridgesensor is a self-contained, AC powered, signal conditioning modul

Strona 2

CF 28 2 4/2001 Specifications (Typical @ 25°C unless noted) Amplifier Gain Range 40-250 Gain Temperature Coefficient 200 ppm/°C Drift (RTI) 3 µV

Strona 3 - CF 28 3 4/2001

CF 28 3 4/2001 C. Turn on the AC source supply to the DCM 460. D. Set the required EXCITATION supply voltage to the load cell by adjusting B+ ADJUST

Strona 4

CF 28 4 4/2001 The sense lines minimize variations in output voltage with changes in load current or lead resistance. It should be noted that if th

Strona 5 - CF 28 5 4/2001

CF 28 5 4/2001 Filter Roll Off Frequency The table in Figure 4 shows the values for C1 and C2 to set the 3 dB filter frequency. A printed circuit bo

Strona 6

CF 28 6 4/2001 Figure 5. Case Dimension Terminal Strip Assignments Screw Terminal Function Screw Terminal Function 1 SENSE 8 +INPUT 2 B- 9